Backflow prevention:
Conductivity:
Chloride content:
Air supply
Pressure range:
Flow rate:
Humidity:
Quality:
Oil contamination:
Particulate
contamination:
Connection:
Water and air filters
Water filter:
Air filter:
Replacement of water and air filters should be performed by qualified
service technician according to the maintenance schedule.
See section "Annual maintenance" on page 6.
Suction connection
Vacuum:
Flow rate:
Connection:
Type:
Class:
Drain connection
Capacity:
Connection:
Output pressure:
The water supply line must be equipped with a
backflow prevention valve according to local
requirements. E.g. in most European countries
according to EN1717 standard, BA Type
backflow preventer.
In dental units with Planmeca WEK water
disinfection system or Planmeca ActiveAqua
water treatment system, a backflow prevention
valve is not needed as they are separated from
mains water with an AB type air gap.
> 80 μS/cm
This requirement applies only to dental units with
Planmeca ActiveAqua water treatment system
Recommended above 8 mg/L but feasible 4 - 8
mg/L.
This requirement applies only to dental units with
Planmeca ActiveAqua water treatment system
min. 550 kPa (80 psi), max. 900 kPa (130 psi)
≥ 55 litres / minute (maximum consumption at
any instance)
dew point not greater than −20 °C at atmospheric
pressure
medical grade, dry and oil-free
3
max. 0.5 mg/m
≤ 100 particles per cubic meter for 1 μm to 5 μm
particle size
1/4"
25 µm (internal), 5 µm (external suggested)
25 µm (internal)
≥ 150 mbar
≥ 550 l/ min
Ø 50 / 46 mm
Medium volume
Wet or dry depending on installed options
min. 10 l/min
Ø 50 / 46 mm
water 2.8 bar, air 5.5 bar (regulated internal water
and air pressure
35 Technical information