How to clean dental unit surfaces
When
After the working
day
Parts that should be cleaned weekly or monthly
When
Weekly
Monthly
Instrument console
User's manual
Part
Cleaning agent
Upholstery
Planmeca
approved
upholstery
disinfectant
On Flexy-holder:
Planmeca
Suction tube
approved surface
holders (incl.
disinfectant
rollers),
instrument holder
and
supplementary
holders
On Flexy-holder:
Planmeca
Holder for
approved surface
intraoral scanner
disinfectant
Part
Upholstery
Instrument flushing
holder
Suction tube
cleaning holder
Coarse filters
Before cleaning the instrument console, lock the touch display by pressing
the Lock button in the
When the touch display is locked, the Lock button is blue. Press the button
for 1 second to unlock the touch display. A progress bar is displayed while
you press the button.
Alternatively, the dental unit can be configured so that you can lock/unlock
the touch display by pressing the Flexy button. Contact your Planmeca
dealer.
Additional cleaning method
Dish-
washer
(65°C)
Cleaning method
Treatment with Dürr FD 360.
Wipe away any excess oil after treatment.
Washer-disinfector (93°C) or autoclave (134°C).
See also section "Instrument flushing holder" on page
241.
Washer-disinfector (93°C).
See also section "Suction tube cleaning holder and
separation tank" on page 241.
Replace them.
See also section "Coarse filters" on page 242.
Maintenance window.
29 Cleaning and disinfection
Washer-
Autoclave
disinfector
(134°C)
(93°C)
X
Planmeca Compact i
237