Help messages
HELP
MEANING
CODE
H 27
Drive backrest upwards, then lower chair. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
H 28
Planmeca Compact WE's help message.
H 29
Drive chair downwards, then lower backrest.
If patient chair is attached to unit, chair's
automatic positions are not switched on.
H 30
Close water tap. Fill container with
disinfectant up to groove and attach to unit.
H 31
Switch unit off, leave disinfectant in unit
overnight (min. 8 h).
H31.1
Remove container, fill it up to groove with
water and attach to unit.
H31.2
Attach container filled with water.
H 32
Container flushing completed.
H 33
Open water tap.
H 34
No instruments selected when starting
instrument flushing or waterline cleaning.
H 35
Instruments are locked for safety reasons.
H35.1
Waterline cleaning cancelled or failed.
Remove container.
H 36
Flushing ready. Return instruments to
instrument console.
H36.1
Instrument flushing completed, but water
flow was not detected for one or more
instruments.
H36.8
Low water pressure during final flushing.
H 37
Waterline cleaning has been interrupted.
H 38
Low water pressure during final rinsing.
User's manual
32 Help and error messages
ACTION OR EXPLANATION
Drive the backrest upwards before lowering
the chair. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair's automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the "AUTO" position (switch up).
Connect the lifter cable.
Drive the chair downwards before lowering
the backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair's automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the "AUTO" position (switch up).
When unit is turned back on, disinfection
procedure continues.
Remove container from unit.
Select at least one instrument and syringe
(remove from holder, bend the instrument
arms).
Select instruments, place them in flushing
holder, and press Instrument flushing button
to start instrument flushing.
Make sure water tap is open. Select
instruments, place them in flushing holder
and press Instrument flushing button to start
instrument flushing.
Also bend and release syringe's instrument
arm once.
Run instruments for a while with spray water
to ensure instrument flushing is sufficient.
Perform long instrument flushing to ensure
instruments are flushed.
When waterline cleaning is finished, perform
long instrument flushing to ensure
instruments are flushed.
Planmeca Compact i
271