32 Help and error messages
280
Planmeca Compact i
Power supply related error messages
ERROR
EXPLANATION
CODE
E7
Power supply related error messages. IPS voltage failure on
MCB.
E9
Internal operating voltage too low. Unit functionality may be
limited.
E10
Internal operating voltage too high.
E11.1
OP-light fuse (F5) and/or Water heater fuse (F7) has blown on
the MCB.
E11.2
Separator fuse (F6) and/or Syringe heater fuse (F8) has blown
on the MCB.
E11.3
Either of the +24V electronics fuses (F9 or F10) has blown on
the MCB.
E11.4
Rectifier D5 or D1 short circuit or either of the 24V electronics
fuses (F3 or F4) has blown on the MCB. Chair, instrument and
headrest control not allowed.
E11.5
Separator fuse (F5) has blown on the MCB.
E11.6
Water heater fuse (F6) has blown on the MCB.
E11.7
Syringe heater fuse (F7) has blown on the MCB.
E11.8
Polymerisation light fuse (F8) has blown on the MCB.
E12.1
SELV voltage too low. Unit functionality may be limited.
E12.2
SELV voltage too high.
E13
Mains frequency outside acceptable range.
E14
IPS temperature measurement sensor error.
E15
IPS heatsink is running too hot on MCB.
E16
Attached scaler electronics is of new type.
E16.1
New type of scaler electronics is attached to MCB.
Update MCB software.
E16.2
New type of scaler electronics is attached to IMUX.
Update MCB software.
E16.3
Scaler electronics slot does not match with scaler hose
connection.
Check scaler electronics connection.
E16.4
Other instrument than scaler is connected to slot where scaler
electronics is.
Check scaler electronics connection.
E17
Power Supply related error messages.
E18
Power Supply related error messages.