Help messages
HELP
MEANING
CODE
H 52
Patient chair can not be driven upward
because safety switch of side delivery arm
is released.
H 53
Patient chair can not be driven downward
because safety switch of side delivery arm
is released.
H 54
Cup is inserted in cup holder.
H 55
Disinfectant container attached. To start
waterline cleaning, select instruments and
press Waterline cleaning button.
H 57
Vision action failed. Select patient in
Romexis.
H 60
WMS: Container almost empty in Domestic
water supply -mode.
H 61
WMS: Container almost empty in Bottled
water -mode.
H 62
WMS: Container switch is turned off.
H 63
Cuspidor door is open.
H 64
Fill container with disinfectant over upper
sensor.
H 65
Fill container with water over upper sensor.
H 66
Wait 8 hours or switch unit off for the night.
H 67
Instrument flush cancelled. Return
instruments to instrument console.
H 68
Romexis connection lost.
H 69
Unable to create a new user. Maximum
amount of local users exceeded.
H 70
Brushless micromotor is stuck or broken.
H 71
Romexis user ID already in use.
H 72
Incompatible Romexis version.
H 72.1
Incompatible Romexis version
H 73
Failed to copy user profile from Romexis.
H 73.1
Failed to save user settings.
H 74
PlanID reader missing.
User's manual
32 Help and error messages
ACTION OR EXPLANATION
Remove the cup and make sure the cup fill
tube is above the bowl.
Select patient in Planmeca Romexis before
performing Solanna Vision action.
Reduce water use for a moment.
Reduce water use and fill the container as
soon as possible.
Switch on the container switch.
Close the door.
Make sure the Ethernet cable is attached
and Romexis is running.
Check the ID, select another one, or allow
rebind in Romexis.
PlanID is not supported with this version of
Romexis.
User settings for currently active instrument
can not be used. Romexis version does not
support user instrument settings or presets
for currently active instrument. Check the
instrument settings before using instrument
and update Romexis.
PlanID reader is either missing or not
functional.