Planmeca Compact i5 User Manual page 235

Table of Contents

Advertisement

User's manual
1 Luzzani Ergo syringe
Remove the metallic syringe cover and place the syringe in the
flushing holder. Make sure you insert the syringe the right way so
that it fits into the holder.
2 DCI syringe
Remove the syringe cover and place the syringe in the flushing
holder. Lock the syringe into place with a rotating movement when
the syringe is in the holder.
3 Luzzani Minibright syringe
Remove the metallic syringe cover and place the syringe in the
flushing holder. Lock the syringe into place by pushing it
downwards until you hear a click.
5. Start the cleaning cycle.
Balanced instrument arms: Start the cleaning cycle by bending the
instrument arms of those instruments (including syringe) that are to be
cleaned to an angle of at least 90° and simultaneously selecting
Waterline cleaning on the control panel.
Hanging-tube instruments: Start the cleaning cycle by selecting
Waterline cleaning on the control panel.
28 Flushing and cleaning programs
Planmeca Compact i
225

Advertisement

Table of Contents
loading

Table of Contents