28 Flushing and cleaning programs
The instrument hose positions on the control panel correspond to their
positions on the instrument console. A selected item is displayed in blue,
and an item that is being cleaned is displayed in a blinking green. When all
items have been cleaned, they are displayed in a steady green. If the
cleaning procedure for some reason is interrupted or not successful (for
example, there is no water flow), the item is displayed in yellow.
To enter the Maintenance mode, press the Maintenance button.
Maintenance window shows the different cleaning programs.
The
The time stamp under the cleaning program name is an easy way to check
when the cleaning program was last started. However, it does not tell
whether the cleaning program was successfully completed.
Suction cleaning also contains information on how many suction cleaning
cycles can be run before filling the suction disinfectant container. When you
have filled the container, you must reset the counter, see section
"Confirming suction disinfectant container fill" on page 200.
While performing the flushing/cleaning, follow the instructions displayed on
the control panel. When the cleaning program runs without problems, the
messages only show briefly, but all messages can be viewed in the
message history log.
To view the message history log, see section "Viewing help and error
message history" on page 290.
Some control panel buttons are common for all cleaning programs, see the
following table.