32 Help and error messages
Help messages
HELP
MEANING
CODE
H 16
Operating light is dimmed and
polymerisation light can be used.
H 17
No syringe connected to instrument
console.
H 18
No instrument hose connected to this
instrument place.
H 19
No syringe connected to connector located
on cuspidor.
H 20
Wrong instrument is connected to cuspidor
syringe connector or assistant instrument
configuration is wrong.
H 21
Flow of both spray air and spray water are
programmed to zero (no flow).
H 22
Main air or water valve is set to be always
closed in SM 12 (air) /13 (water).
H 23
Automatic/manual chip blow cannot be used
when sterile water function is selected.
H 24
Programming of selected setting is disabled. The programming of the selected setting is
H 25
Planmeca Compact e help message:
Patient chair cannot be driven into
automatic chair position.
H 26
Drive chair upwards, then lower backrest. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
270
Planmeca Compact i
ACTION OR EXPLANATION
This feature enables easier work with
composite materials, minimising the risk of
operating light induced curing. This feature
is triggered by picking up and returning the
polymerisation light to its holder without
starting it. The operating light returns to
normal at the same instance any control is
touched or another instrument selected. The
feature can be disabled.
Contact your Planmeca dealer.
The console syringe must be re-connected,
otherwise there is an air-leak from the
syringe QC whenever an instrument or
device that requires air pressure is used.
The instrument is picked up, but there is no
hose connected at this location. Check that
all hoses go to the right instrument arms.
The assistant's syringe must be re-
connected, otherwise there is an air-leak
from the assistant's syringe QC whenever
an instrument or device that requires air
pressure is used.
Contact your Planmeca dealer.
Contact your Planmeca dealer.
Contact your Planmeca dealer.
This is a reminder that either of the main
valves has been set to be always closed.
The setting may be changed in service
modes SM 12 (air) and SM 13 (water).
disabled.
Contact your Planmeca dealer.
Contact your Planmeca dealer.
Drive the chair upwards before lowering the
backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair's automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the "AUTO" position (switch up).