31 Clean water system
Results
266
Planmeca Compact i
5. Start the short flushing cycle.
Balanced instrument arms: Start the flushing cycle by bending the
instrument arms of those instruments (including syringe) that are to be
flushed to an angle of at least 90° and simultaneously selecting Short
flushing on the control panel.
Hanging-tube instruments: Start the flushing cycle by selecting Short
flushing on the control panel.
The system will first identify the instruments and then each instrument
hose is flushed with air and water for 30 seconds in its turn.
6. Release the instrument arms (if your dental unit is equipped with
balanced instrument arms).
7. Repeat the flushing sequence in steps 5 and 6 at least three more times,
or until the water is clean. If the clean-water bottle runs out of water, refill
it.
If the water quality of the domestic water is good, you can also use the
domestic water mode when performing the flushing.
8. Balanced instrument arms: Place the instrument hoses back in the
instrument arms, and the instruments back in the instrument console.
Hanging-tube instruments: Place the instruments back in the instrument
holders.
The dental unit is now ready for use.