Planmeca Compact i5 User Manual page 312

Table of Contents

Advertisement

35 Technical information
302
Planmeca Compact i
F1, F2 = Schurter 0001.1014 10A/250V/FAST ACTING/HIGH BR CAP.
(100V, 115V)
F1, F2 = Bussmann S501-10-R 10A 250V FAST ACTING/HIGH BR CAP.
(100V, 115V)
F1, F2 = Schurter 0001.1012 6.3A/250V/FAST ACTING/HIGH BR CAP.
(220V – 240V)
F1, F2 = Bussmann S501-6.3-R 6.3A 250V FAST ACTING/HIGH BR CAP.
(220V- 240V)
F3, F4 = 4A/250V/Fast act./High br. cap, Schurter 0001.1010 OR
Bussmann S501-4-R
F5 = 8A/250V/Fast act./High br. cap, Schurter 0001.1013 OR Bussmann
S501-8-R
F6 - F8 = 6.3A/250V/Fast act./High br. cap, Schurter 0001.1012 OR
Bussmann S501-6.3-R
Power consumption
Idle unit:
Typical average:
Maximum:
Electrical classification
Class I
Operation of chair lift and backrest motors
Intermittent operation, ED 6%, 25 sec "ON", 400 sec. "OFF"
Water supply
Pressure range:
Flow rate:
Quality:
pH:
Connection:
≈ 60 VA (unit not in use, OP-light turned on)
≈ 350 VA (during patient treatment)
1450 VA (at maximum rated mains voltage, both
chair motors running)
min. 300kPa (44 psi), max. 900 kPa (130 psi)
In dental units with Planmeca WEK water
disinfection system, the pressure range must be
300 - 600 kPa (44 – 87 psi).
In dental units with Planmeca ActiveAqua water
treatment system, the pressure range must be
(100)200 - 900 kPa ((15)30 - 130 psi).
≥ 4 l / min (maximum consumption at any
instance)
hardness; ≤ 8°dH
In dental units with Planmeca ActiveAqua water
treatment system, the water hardness can be 0 -
40°dH.
6.5 – 8.5