28 Flushing and cleaning programs
Automatic feeding begins and the instrument waterlines, the cup fill line
Waterline cleaning window on the
and the bowl waterlines are filled. The
control panel informs you about the progress of the cleaning cycle.
11. When the disinfectant has been fed, help message H 31 is displayed
(you will hear a signal tone).
This means that you should now turn off the unit's power.
Let the disinfectant stay in the unit overnight (min 8h). After the waiting
period you may proceed to section "Waterline flushing" on page 230.
NOTE
The disinfectant must not be left in the unit for longer than one night (not,
for example, over the weekend).
User's manual
Planmeca Compact i
229