32 Help and error messages
Help messages
HELP
MEANING
CODE
H 75
Yearly maintenance in <#> days.
H 76
Yearly maintenance performed.
H 79.1
Serial number chip missing.
H 80
Headrest length safety feature activated
when driving headrest. Check that nothing is
blocking headrest movements.
H80.1
Headrest length drive limit activated when
driving headrest.
H 81
Headrest angle B safety feature activated
when driving headrest. Check that nothing is
blocking headrest movements.
H81.1
Headrest angle A drive limit activated when
driving headrest.
H 82
Connecting to wireless foot control was
successful.
H 83
Foot control handle pressed down.
H 84
Connecting to wireless foot control failed.
H 85
Software update denied. Chair was moving
or instrument was in use while starting
software update.
H 86
Software update denied. Incorrect control
panel type was detected while starting
software update.
H 87
Software update halted. Activate the
wireless foot control by pressing the handle.
H 88
Wireless foot control configuration was
successful.
H 89
Wireless foot control configuration failed.
274
Planmeca Compact i
ACTION OR EXPLANATION
Contact service for yearly maintenance.
The serial number chip is missing. Contact
service.
Check that nothing is blocking the headrest
movement. The headrest can be driven
normally after the possible obstruction has
been removed.
Check that nothing is blocking headrest
movements.
Faulty length motor, motor cable,
nonlubricated slider rail or position
potentiometer may cause this problem.
Check that nothing is blocking the headrest
movement. The headrest can be driven
normally after the possible obstruction has
been removed.
Faulty angle B motor, motor cable or
position sensor may cause this problem.
Check that nothing is blocking the headrest
movement.
Faulty angle A motor, motor cable or
position sensor may cause this problem.
Connection was established with the
requested foot control.
Release the foot control handle.
Connection with the requested foot control
could not be established. Make sure the foot
control is active (connect to a charger, if
possible), is in the vicinity of the unit, and try
again.
The chair was moving or an instrument was
in use while starting the software update.
An incorrect control panel type was detected
while starting the software update.
Wireless foot control has to be activated
before it can be updated.
Wireless foot control channel and power
settings were successfully changed.
Wireless foot control channel and power
settings could not be changed. Make sure
the foot control is active (connect to a
charger, if possible), is in the vicinity of the
unit, and try again.