Planmeca Compact i5 User Manual page 279

Table of Contents

Advertisement

Help messages
HELP
MEANING
CODE
H 3
Patient chair cannot be driven down
because safety switch is activated.
H 4
This left-hand instrument position allows
only syringe.
H 5
This instrument position does not accept
syringe.
H 6
Patient chair cannot be driven when
instrument is in use.
H 7
Patient chair up/down movement is not
allowed when cuspidor door is open.
H 8
Chair height can not be programmed this
high in automatic positions for patient safety
reasons.
H 9
This scaler cannot currently be used with
this unit.
H 10
Wait while configuring the dental instrument
setup.
H 11
Software download mode is on.
H 12
Unit's internal error reporting is disabled.
H 13
Instrument cannot run because of patient
security violation.
H 14
This dental instrument is not recognised
(new type of instrument).
H 14.1
Instrument presets reset to defaults.
H 15
This button has no function at the moment.
User's manual
32 Help and error messages
ACTION OR EXPLANATION
Check that the space under the seat is free
and nothing is touching the plate at the
bottom. The chair can be driven normally
after the possible obstruction has been
removed.
The leftmost position in the console is
reserved for the syringe only. Replace this
instrument with a syringe. Place the other
instrument to one of the four right-hand
places.
Remove the syringe from this position and
replace it with any other kind of instrument.
The syringe must be placed in the leftmost
place.
Stop the instrument (release the foot control
pedal) to be able to run the chair.
Close the cuspidor door to be able to run
the chair up/down or to a pre-programmed
position.
Run the chair slightly down to be able to
program this position into memory.
This scaler needs a dedicated electronics
that is not installed, or the scaler electronics
installed is not compatible with this scaler.
Wait a few seconds for the unit to check
your instrument configuration, since it has
changed.
The unit operates normally but no errors are
reported (exhibition use). Use service mode
n.15 to turn error reporting back on.
Return the foot control pedal once to the
middle position to start the instrument.
Instruments do not start if the standard foot
control pedal is already pushed to the right
or left when the instrument is picked up from
the instrument console.
To be able to operate this new instrument
the MCB software must be updated.
Contact your Planmeca dealer.
Instrument presets are faulty. Presets are
reset to defaults.
Instruments: Pick up a dental instrument
before attempting to alter its settings; or if
the instrument is selected, this function is
not allowed with this instrument. Other
buttons: the unit does not have this function.
Planmeca Compact i
269