25.4.9 Planmeca Lumion Plus polymerisation light
About this task
Steps
25.5 Timer settings
About this task
Steps
User's manual
NOTE
The duration of the polymerisation cycle can be programmed only for the
Planmeca Lumion Plus polymerisation light.
1. Activate the instrument.
2. Press Program.
3. Press Polymerisation cycle.
A pop-up window opens.
4. Adjust the length of the polymerisation cycle.
The value range is 5 - 100 seconds. The adjustment step is 5 sec. and
the default value is 10 sec.
5. Press OK.
The pop-up window closes and the new values are displayed on the
control panel.
Up to six timer settings can be programmed into the timer memory.
1. Press Program.
2. Press Timer.
A pop-up window opens.
3. Press the timer that you want to adjust.
A new pop-up window opens.
4. Adjust the length of the timer.
The minimum value is 5 sec. and the maximum 20 min. The adjustment
step is 5 sec.
25 Programming
Planmeca Compact i
191