NPD checkout conditions

Getting Started
Set up the GC as shown in
of setpoint entry.
Table 16
NPD checkout conditions
Column
Sample
Inlet
Temperature
Initial pressure
Mode (split/splitless)
Purge time (split/splitless)
Purge flow (split/splitless)
Detector
Temperature
Makeup + column flow
Offset
Oven
Init temp
Init time
Rate
Final temp
Final time
Checking Performance
Table
16. See
page 79
HP-5MS, 30 m x 0.32 mm x 0.25 µm,
part number 19091S-413
1 µL of NPD checkout, 18789-60060
Split/splitless
200 °C
0.172 MPa (25 psi) for 7 mL/min
at 60 °C
Splitless
0.75 min
60 mL/min
NPD
300 °C
10 mL/min (nitrogen recommended)
30 pA
60 °C
0 min
20 °C/min
200 °C
4 min
for an example