Swagelok Connections
96
Pencil line
Figure 51
Marking the fitting
8 For 1/8-inch Swagelok fittings, use a pair of 7/16-inch
wrenches to tighten the fitting 3/4 of a turn
For 1/4-inch fittings, use a pair of 9/16-inch wrenches to
tighten them 1-1/4 turn
Figure 52
Final tightening
9 Remove the plug from the fitting. To connect the tubing, with
nut and ferrules, to another fitting, finger-tighten the nut,
then use a wrench to tighten it 3/4 (1/8-inch fittings) or 1-1/4
(1/4-inch fittings) of a turn.
(Figure
52).
(Figure
52).