The 6820 Gas Chromatograph
WA R N I N G
16
It is unlikely, even in this very unusual situation, that
radioactive material will escape the cell. However, permanent
63
damage to the
Ni plating within the cell is possible, and
therefore, the cell must be returned for exchange.
Do not use solvents to clean the ECD.
You may not open the ECD cell unless authorized to do so by your
local nuclear regulatory agency. Do not disturb the four socket-head
bolts. These hold the cell halves together. Removing or disturbing
them is a violation of the terms of the General License and could
create a safety hazard.

Safety precautions when handling ECDs

Always observe the following precautions:
• Never eat, drink, or smoke when handling ECDs.
• Always wear safety glasses when working with or near open
ECDs.
• Wear protective clothing such as laboratory jackets, safety
glasses, and gloves, and follow good laboratory practices.
Wash hands thoroughly with a mild non-abrasive cleaner
after handling ECDs.
• Cap the inlet and outlet fittings when the ECD is not in use.
• Connect the ECD exhaust vent to a fume hood or vent it to
the outside.
Agilent Technologies recommends a vent line inside diameter
of 6 mm (1/4-inch) or greater. With a line of this diameter,
the length is not critical.
Getting Started