3 Push the tubing into the stainless steel plug
4 Make sure that the front ferrule is touching the plug. Slide
the Swagelok nut over the ferrule and thread it onto the plug.
Nut and ferrules
Plug or fitting
held in a vise
Figure 49
Assembling the fitting
5 Push the tube fully into the plug, then withdraw it
approximately 1 to 2 mm
Front
ferrule
Figure 50
Insert the tubing
6 Finger-tighten the nut.
7 Mark the nut with a pencil line
Swagelok Connections
(Figure
(Figure
50).
Nut
Insert tubing fully
Back
ferrule
Withdraw tubing 1—2 mm
Tighten nut
(Figure
51).
49).
Tubing