8.3 Wiring of the Outputs
Option 1: The gaussmeter switches the positive potential provided by the customer.
The load is switched against the external minus pole.
This option is preferably used in European countries in order to connect a PLC
control.
Option 2: The gaussmeter switches the minus potential provided by the customer. The
load is switched against the external plus pole.
This option is preferably used in America and Japan in order to connect a PLC
control.
Page 44 / 106
Insulated DC output
Used as current source outputs.
Special features.
Do not connect!
Figure 17 Signal Output Plus Switched
Insulated DC output
Used as current sinking outputs
Special features
Do not connect!
Figure 18 Signal Output Minus Switched
Gaussmeter IGM11 Operating Instructions