MAGSYS IGM11 Operating Instructions Manual page 32

Industrial gaussmeter
Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

magne t
MAGSYS
systeme
The error of the fast peak value measurement consists of the error of the DC field
measurement and of a frequency factor that depends on the measuring range.
The range-depending frequency response factor is multiplied by the basic accuracy,
depending on the respective signal frequencies and then added to the basic
accuracy.
Example:
Measurement in the range of 1000mT with a frequency of 1kHz.
The basic accuracy lies at 0.5%. The frequency response factor is 0.98 = -2%.
During this measurement, you have to calculate a total error of -2%±0.5%.
Page 32 / 106
Error DC Field Measurement
B ≤ 1.5 T
B ≥ 1.5 T
Range-Depending Frequency Response Factor (Sinusoidal Field Pattern)
Frequency
4500 mT
50 Hz
1.00
1.00
70 Hz
1.00
100 Hz
1.00
200 Hz
500 Hz
0.99
700 Hz
0.99
1000 Hz
0.98
2000 Hz
0.92
Gaussmeter IGM11 Operating Instructions
(1σ)
≤ ±0.5 %
≤ ±1.0 %
Measuring Range
1000 mT
100 mT
1.00
1.00
1.00
0.99
1.00
0.98
1.00
0.99
0.98
0.98
Chapter 5 Operation
10 mT
0.99
0.98
0.93