EIA232 buffer overflow
EIA485 parity error
EIA485 framing error
EIA485 overrun error
EIA485 buffer overflow
EIA485 checksum error
EIA485 adress error
Undefined error
Gaussmeter IGM11 Operating Instructions
Chapter 10 Table of Error Messages
Buffer overflow on the serial interface EIA-232.
If necessary, reduce the bit rate. Check the
command.
Character parity error on the serial interface
EIA-485.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
Character framing error on the serial interface
EIA-485.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
Character overrun error on the serial interface
EIA-485.
Possibly, reduce the bit rate.
Buffer overflow on the serial interface EIA-485.
Possibly, reduce the bit rate. Check the
command.
BCC character error on the serial interface.
EIA485.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
Error in the address structure on the serial
interface EIA485.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
Undefined error.
Forward the error number to the service.