Wiring Of The Outputs - MAGSYS IGM11 Operating Instructions Manual

Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

Industrial Gaussmeter IGM11 Operating Instructions
Chapter 8 Signal Interfaces

Wiring of the Outputs

Option 1: The gaussmeter switches the positive potential provided by the customer.
The load is switched against the external minus pole.
This option is preferably used in European countries in order to connect a PLC control.
Option 2: The gaussmeter switches the minus potential provided by the customer. The
load is switched against the external plus pole.
This option is preferably used in America and Japan in order to connect a PLC control.
Page 43 / 92
Insulated DC output
Used as current source outputs.
Special features.
Do not connect!
Figure 17 Signal Output Plus Switched
Insulated DC output
Used as current sinking outputs
Special features
Do not connect!
Figure 18 Signal Output Minus Switched