magne t
MAGSYS
systeme
9.12.1.4 *ESR[?]
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.12.1.5 *IDN?
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.12.1.6 *OPC?
Description:
Mode:
Parameter:
Response:
*RST
Example:
Page 72 / 106
Standard Event Status Enable Query
Reset or query the standard event release register. The gaussmeter
returns the bit pattern of the register as a decimal value.
Command and query
{<Value>} within the range of 0..255
Standard event release register
Value:
not relevant
send
*ESR 22<LF>
send
*ESR?<LF>
receive
160<CR><LF>
Identification Query
Query the gaussmeter identification character string. The gaussmeter
returns the following identification text:
MAGSYS-MAGNET-SYSTEME,IGM11,1,xxxxxx,C
Query
none
Identification character string
Value:
not relevant
send
*IDN?<LF>
receive
MAGSYS-MAGNET-SYSTEME,IGM11,1,011111,C<CR><LF>
Operation Complete Query
Return of "1" to the output buffer, after the command has been
executed. The command *OPC? can be placed at the end of a
command line in order to synchronize the controller with the
command processing. The command itself has no function.
Query
none
1
Value:
not relevant
send
*OPC?<LF>
receive
1<CR><LF>
Gaussmeter IGM11 Operating Instructions