Industrial Gaussmeter IGM11 Operating Instructions
Chapter 10 Table of Error Messages
10 Table of Error Messages
Errors inside the device are stored in the error queue. It can be read out via an SCPI
command or via the set-up menu.
The messages have the following meanings:
Error
Number
0
-101
-102
-113
-121
-151
-222
-224
-350
522
100
101
102
103
104
105
200
201
202
Error Text
No error
Invalid character
Syntax error
Undefined header
Invalid character in number
Invalid string data
Data out of range
Illegal parameter value
Too many errors
Output buffer overflow
Param. memory error
Probe parameter error
Internal parameter error
Internal error WD
Internal error ADRR
Internal error STCK
EIA232 parity error
EIA232 framing error
EIA232 overrun error
MAGSYS
Reason / Solution
There is no error.
SCPI error: An invalid character has been
detected. Check the command.
SCPI error: An invalid parameter has been
detected. Check the command.
SCPI error: An invalid command has been
detected.
Check the command.
SCPI error: An invalid character in a number
has been detected. Check the command.
SCPI error: An invalid character in the text has
been detected. Check the command.
SCPI error: Numerical value exceeds the
allowed limits.
Check the command.
SCPI error: Invalid parameter has been
detected. Check the command.
Error queue is full.
Output buffer for interface is too small. Check
the command.
Parameter memory data are inconsistent.
Check the parameter settings.
Calibration data of the measuring probe
cannot be queried by the device.
Re-connect the probe or try another probe.
Internal calibration data are inconsistent.
Switch the device off and on.
An internal error has occurred.
Switch the device off and on.
Character parity error on the serial interface
EIA-232.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
Character frame error on the serial interface
EIA-232.
Check the settings of your controller and of the
device. Possibly, the transmission path is
disturbed.
Character overrun error on the serial interface
EIA-232.
If necessary, reduce the bit rate.