MAGSYS IGM11 Operating Instructions Manual page 100

Industrial gaussmeter
Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

magne t
MAGSYS
systeme
Measuring Properties
Display refresh time
Display resolution
Frequency range
Meas. inaccuracy
Page 100 / 106
Logic 1 (min.)
20 V DC with maximum current
Logic 0 (max.)
0.1 V DC with 10 k
Function
Individually selectable via parametrization
Output shunt
Possible
Connector
Via common12-pole IO connector; spring force connection; Nominal
diameter 0.14 to 0.5 mm²; AWG 28 to 20; stripping length 8 mm;
Connection according to norm EN-VDE; test points;
Connection of the common conductor via the current supply
connector.
Meas. method
Continuous recording of the magnetic field; conversion via 16-bit A/D-
converter; Evaluation via 16 bit microprocessor system.
100 ms;
Depending on range 3 to 4-digit
DC / AC 0 Hz .. 5 kHz (effective value)
Accuracy (1σ)
DC-field-
measuremen
t
AC-field
measuremen
t
Peak value
measuremen
t
Gaussmeter IGM11 Operating Instructions
Chapter 12 Technical Data
load
B
≤ 1.5 T
> 1.5 T
Frequency
B
eff
≤ 2 kHz
≤ 1 T
≤ 5 kHz
≤ 2 T
Range
≤ ±1.0 %
10 mT
≤ 70 Hz
100 mT
≤ 100 Hz
1 T
≤ 300 Hz
4.5 T
≤ 500 Hz
(B < 1.5 T)
Error
≤ ±0.5 %
≤ ±1.0 %
≤ ±1.0 %
≤ ±2.0 %
≤ ±2.0 %
≤ 100 Hz
≤ 150 Hz
≤ 500 Hz
≤ 700 Hz