MAGSYS IGM11 Operating Instructions Manual page 92

Industrial gaussmeter
Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

magne t
MAGSYS
systeme
9.12.7.14 :DGIO:SET:OUT
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.12.7.15 :DGIO:SET:OUT
Description:
Mode:
Parameter:
Response:
*RST
Example:
Page 92 / 106
Set or query the function of each output.
x is for outputs 1 to 4.
Command or query
{
NO|IFHI|IFLO|IFIN|IFOU|IFOV|IFRN
Parameter
NO
IFHI
IFLO
IFIN
IFOU
IFOV
IFRN
like parameter
Value:
not relevant
send
:DGIO:SET:OUT2:FUNC IFOV<LF>
send
:DGIO:SET:OUT2:FUNC?<LF>
receive
IFOV<CR><LF>
Set or query the logic level of each output.
x is for outputs 1 to 4
Command or query
{
}
0|1
0 is for normal function (current flow when this condition is fulfilled)
1 is for inverting function
like parameter
Value:
not relevant
send
:DGIO:SET:OUT2:LOG 1<LF>
send
:DGIO:SET:OUT2:LOG?<LF>
receive
1<CR><LF>
Gaussmeter IGM11 Operating Instructions
:FUNC[?]
{x}
}
Function
No function assigned to this output
This output is set if the measuring value is geater
than the set upper limit value.
This output is set if the measuring value is less than
the set lower limit value.
This output is set if the measuring value is within the
range of upper and lower limit values.
This output is set if the measuring value exceeds the
range of lower and upper limit values.
This output is set if the measuring range has been
exceeded.
This output is set if the device is ready to measure.
:LOG[?]
{x}
Chapter 9 Serial Interfaces