MAGSYS IGM11 Operating Instructions Manual page 77

Industrial gaussmeter
Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
9.12.2.3 :RANGe:SET
Description:
Mode:
Parameter:
Response:
*RST
Example:
Meaning of the parameters with different units and measuring modes.
DC field measurement
Fast peak value
measurement
AC field measurement
9.12.2.4 :RANGe?
Description:
Mode:
Parameter:
Response:
*RST
Example:
Set the measuring range of the gaussmeter.
Command
{0|1|2|3}
0
Most sensitive range
3
Least sensitive range
none
Value:
3
send
Parameter
3
4.5 / 10T
2
and
1
100 mT
0
10 mT
3
3.0 / 6.5 T
2
1
100 mT
0
10 mT
Query the current measuring range.
Query
None
{0|1|2|3}
0
Most sensitive range
3
Least sensitive range
Value:
not relevant
send
receive
:RANG:SET 2<LF>
Tesla
Gauss
45 / 100 kG
1 T
10 kG
1 kG
100 G
30 / 65 kG
1 T
10 kG
1 kG
100 G
:RANG?<LF>
3<CR><LF>
magne t
MAGSYS
systeme
Oersted
A/m
45 / 100 kOe
3800 / 8500 kA/m
10 kOe
1000 kA/m
1 kOe
100 kA/m
100 Oe
10 kA/m
30 / 65 kOe
2500 / 5500 kA/m
10 kOe
1000 kA/m
1 kOe
100 kA/m
100 Oe
10 kA/m
Page 77 / 106

Advertisement

Table of Contents
loading

Table of Contents