Sampler Magnitude and hase Correction Constants (Test 53)

Required Equipment and Tools
ower meter
ower sensor
ower sensor (for Option 006 analyzers)
ower splitter
RF cable set (2)
H -IB cable
Adapter type-N (m) to type-N (m)
Adapter type-N (f) to type-N (f)
Antistatic wrist strap
Antistatic wrist strap cord
Static-control table mat and earth ground wire
Analyzer warm-up time:
This adjustment procedure corrects the overall atness of the microwave components that
make up the analyzer receiver and test separation sections. This is necessary for the H
8753D Option 011 to meet the published test port atness.
1. If you just completed \Source Correction Constants (Test 47)," continue this procedure
with step 8.
2. ress
PRESET 5 LOCAL 5
3. ress
ET ADDRE E
LOCAL 5
is 13. Refer to the power meter manual as required to observe or change its H -IB
address.
4. ress
POWER MTR:438A/437
Choose the appropriate model number.
ote
If you are using the H 438A power meter, connect the H 8482A power
sensor to channel A, and for Option 006, connect the H 8481A power sensor
to channel B.
3-22 Adjustments and Correction Constants
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
0 minutes.
.
Y TEM CONTROLLER
ADDRE : P MTR/HPIB
to toggle between the 438A/437 and 436A power meters.
H 437B or H 438A
H 11667A Option 001
H
/N 1250-1475
H
/N 1250-1472
H
/N 9300-1367
H
/N 9300-0980
H
/N 9300-0797
. The default power meter address
3/21/106 15:11
H 8482A
H 8481A
H 11500B
H 10833A
RAFT