Front Panel Assembly
7. emove both front feet (item 8).
8. emove the four screws (item 9) from the bottom edge of the frame.
9. Slide the front panel over the four Type-N connectors (item 10).
10. Disconnect the ribbon cable (W17) from the front panel by pressing down and out on the
connector locks. The front panel is now free from the instrument.
Replacement
everse the order of the removal procedure.
Note
When reconnecting semi-rigid cables, it is recommended that the connections
be torqued to 10 in-lb.