Accessories Troubleshooting - HP 8753D Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753D Option 011:
Table of Contents

Advertisement

Accessories Troubleshooting

Use this procedure only if you have read Chapter 4, \Start Troubleshooting Here." Follow the
procedures in the order given, unless instructed otherwise.
Measurement failures can be divided into two categories:
Failures which don't a ect the normal functioning of the analyzer but render incorrect
measurement data.
Failures which impede the normal functioning of the analyzer or prohibit the use of a
feature.
This chapter addresses the rst category of failures which are usually caused by the following:
operator errors
faulty calibration devices or connectors
bad cables or adapters
improper calibration techniques
RF cabling problems within the test set
These failures are checked using the following procedures:
\Inspect the ccessories"
\Inspect the Error Terms"
\Test Set Troubleshooting"
RAFT
3/21/106 15:13
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Accessories Troubleshooting 9-1