Eplacing An Assembly - HP 8753D Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753D Option 011:
Table of Contents

Advertisement

eplacing an assembly

The following steps show the sequence to replace an assembly in an HP 8753D Option 011
Network Analyzer.
1. Identify the faulty group. efer to Chapter 4, \Start Troubleshooting Here." Follow up
with the appropriate troubleshooting chapter that identi es the faulty assembly.
2. Order a replacement assembly. efer to Chapter 13, \ eplaceable Parts."
3. eplace the faulty assembly and determine what adjustments are necessary. efer to
Chapter 14, \Assembly eplacement and Post- epair Procedures."
4. Perform the necessary adjustments. efer to Chapter 3, \Adjustments and Correction
Constants."
5. Perform the necessary performance tests. efer to Chapter 2, \System Veri cation and
Performance Tests."
arning
These servicing instructions are for use by qualified personnel only. To avoid
electrical shock, do not perform any servicing unless you are qualified to do so.
arning
The opening of covers or removal of parts is likely to expose dangerous
voltages. isconnect the instrument from all voltage sources while it is being
opened.
arning
The power cord is connected to internal capacitors that may remain live for
10 seconds after disconnecting the plug from its power supply.
Caution
Many of the assemblies in this instrument are very susceptible to damage from
ESD (electrostatic discharge). Perform the following procedures only at a
static-safe workstation and wear a grounding strap.
14-2 Assembly Replacement and
Post-Repair Procedures
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
RAFT
3/21/106 15:15

Advertisement

Table of Contents
loading