A4 Sampler/Mixer Check
The 4, 5, and 6 (R, and B) sampler/mixers are identical. ny sampler can be used to
phase lock the source. To eliminate the possibility of a bad R sampler, follow this procedure:
1. Connect the power splitter, RF cable and attenuator to inputs (or B) and R as shown in
Figure 7-1.
2. Remove the W8 cable ( 11J1 to 4) from the R-channel sampler ( 4) and connect it to
either the -channel sampler ( 5) or the B-channel sampler ( 6), depending on which one
you selected in step 1. Refer to Figure 7-3.
Figure S MMIX7 here.
Figure 7-3. Sampler/Mixer to Phase Lock Cable Connection iagram
3. If you connected W8 to:
, press
A/R
A5
MEAS
5
, press
B/R
A6
MEAS
5
4. Ignore the displayed trace, but check for phase lock error messages. If the phase lock
problem persists, the R-channel sampler is
7-6 Source Troubleshooting
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
the problem.