RF Output ower Correction Constants (Test 47)

Required Equipment and Tools
ower meter
ower sensor
ower sensor (for Option 006 analyzers)
ower splitter (2)
Attenuator 20 dB
H -IB cable
RF cable 24-inch
Antistatic wrist strap
Antistatic wrist strap cord
Static-control table mat and earth ground wire
Analyzer warm-up Time:
This procedure adjusts several correction constants that can improve the output power level
accuracy of the internal source. They are related to the power level, power slope, power slope
o set, and the ALC roll-o factors among others.
1. If you just completed \Sampler Magnitude and hase Correction Constants (Test 53),"
continue this procedure with step 8.
2. ress
PRESET 5 LOCAL 5
3. ress
ET ADDRE E
LOCAL 5
is 13. Refer to the power meter manual as required to observe or change its H -IB
address.
4. ress
POWER MTR:438A/437
Choose the appropriate model number.
ote
If you are using the H 438A power meter, connect the H 8482A power
sensor to channel A, and the H 8481A power sensor to channel B.
RAFT
3/21/106 15:11
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
0 minutes.
.
Y TEM CONTROLLER
ADDRE : P MTR/HPIB
to toggle between the 438A/437 and 436A power meters.
H 437B or H 438A
H 11667A Option 001
H 8491A Option 020
. The default power meter address
Adjustments and Correction Constants 3-9
H 8482A
H 8481A
H 10833A
H 11500B
H
/N 9300-1367
H
/N 9300-0980
H
/N 9300-0797