14 fractional-N digital
19 graphics processor (disconnect W14, 18W1, and W20)
4. Switch on the analyzer and observe the green LEDs on 8.
If any of the green LEDs are o or
ashing, it is not likely that any of the assemblies
listed above is causing the problem. Continue to \Brie
y Disable the Shutdown
Circuitry."
If all green LEDs are now on, one or more of the above assemblies may be faulty.
Continue to next step.
5. Switch o the analyzer.
6. Reconnect W14 and W20 to 19.
7. Switch on the analyzer and observe the LEDs.
If the LEDs are o or blinking, replace the 19 assembly.
If the LEDs are still on, continue to next step.
8. Switch o the analyzer.
9. Reconnect 18W1 to the 19 assembly.
10. Switch on the analyzer and observe the LEDs.
If the LEDs are o, replace the 18 display.
If the LEDs are still on, continue with the next step.
11. Switch o the analyzer.
12. Reinstall each assembly one at a time. Switch on the analyzer after each assembly is
installed. The assembly that causes the green LEDs to go o or
ash could be faulty.
It is possible, however, that this condition is caused by the 8 post regulator
Note
not supplying enough current. To check this, reinstall the assemblies in a
dierent order to change the loading. If the same assembly appears to be
faulty, replace that assembly. If a dierent assembly appears faulty, 8 is most
likely faulty (unless both of the other assemblies are faulty).
Briefly Disable the Shutdown ircuitry
In this step, you shutdown the protective circuitry is disabled for a short time, and the
supplies are forced on (including shorted supplies) with a 100% duty cycle.
Damage to components or to circuit traces may occur if 8TP4 (SDIS) is
aution
shorted to chassis ground for more than a few seconds while supplies are
shorted.
1. Connect 8TP4 (SDIS) to chassis ground with a jumper wire.
2. Switch on the analyzer and note the signal mnemonics and test points of any LEDs that
are o. Immediately remove the jumper wire .
3. Refer to the block diagram (Figure 5-8) at the end of this chapter and do the following:
Note the mnemonics of any additional signals that may connect to any 8 test point that
showed a fault in the previous step.
5-12 Power Supply Troubleshooting
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
RAFT
3/21/106 15:12