A17 Motherboard Assembly
11. emove the display (refer to \A18 Display" in this chapter).
12. emove the source assembly (refer to \A3 Source Assembly" in this chapter).
13. emove the samplers and pulse generator (refer to \A4, A5, A6 Samplers and A7 Pulse
Generator" in this chapter).
14. emove the card cage boards (refer to \A8, A10, A11, A12, A13, A14 Card Cage Boards"
in this chapter).
15. emove the bezel support (item 3) by removing the three screws (item 4) from the left
side of the front frame.
16. emove the actuator assembly (item 5) and insulator strip (item 6) by unhooking the
spring (item 7) and removing the screw (item 8) that secures them to the display housing.
17. emove the front frame (item 9) and rear frame (item 10) by removing the attaching
screws (item 11). At this point, only the motherboard/card cage assembly should remain.
This whole assembly is replaceable.
Replacement
everse the order of the removal procedure.
14-40 Assembly Replacement and
Post-Repair Procedures