HP 8753D Option 011 with Option 1D5 Only
8. Connect the BNC to BNC jumper between the \EXT REF" and the \10 MHz recision
Reference" as shown in Figure 3-20.
9. Use a
at-head screwdriver to remove the screw that covers the precision frequency
adjustment as shown in Figure 3-20. Insert a narrow screwdriver and adjust the precision
frequency reference potentiometer for a spectrum analyzer center frequency measurement
of 5 Hz.
Figure O T1D5 here.
Figure 3-20. High Stability Frequency Adjustment Location
10. Replace the A26 board assembly if you cannot adjust for a center frequency measurement
of 50 Hz.