HP 8753D Option 011 with Option 1D5 Only
8. Connect the BNC to BNC jumper between the \EXT REF" and the \10 MHz recision
Reference" as shown in Figure 3-20.
9. Use a at-head screwdriver to remove the screw that covers the precision frequency
adjustment as shown in Figure 3-20. Insert a narrow screwdriver and adjust the precision
frequency reference potentiometer for a spectrum analyzer center frequency measurement
of 5 Hz.
Figure O T1D5 here.
Figure 3-20. High Stability Frequency Adjustment Location
10. Replace the A26 board assembly if you cannot adjust for a center frequency measurement
of 50 Hz.
RAFT
3/21/106 15:11
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Adjustments and Correction Constants 3-47