HP 8753D Option 011 Service Manual page 283

Network analyzer
Hide thumbs Also See for 8753D Option 011:
Table of Contents

Advertisement

DISK MESSAGE LENG H ERROR
Error Number
The analyzer and the external disk drive aren't communicating
190
properly. Check the HP-IB connection and then try substituting
another disk drive to isolate the problem instrument.
DISK: not on, not connected, wrong addrs
Error Number
The disk cannot be accessed by the analyzer. Verify power to the
38
disk drive, and check the HP-IB connection between the analyzer
and the disk drive. Ensure that the disk drive address recognized by
the analyzer matches the HP-IB address set on the disk drive itself.
DISK READ/WRI E ERROR
Error Number
There may be a problem with your disk. Try a new oppy disk. If a
189
new oppy disk does not eliminate the error, suspect hardware
problems.
EXCEEDED 7 S ANDARDS PER CLASS
Error Number
When modifying calibration kits, you can de ne a maximum of
72
seven standards for any class.
INI IALIZA ION FAILED
Error Number
The disk initialization failed, probably because the disk is damaged.
47
INSUFFICIEN MEMORY
Error Number
Your last front panel or HP-IB request could not be implemented
51
due to insucient memory space. In some cases, this is a fatal error
from which you can escape only by presetting the instrument.
10-44 Service Key Menus and Error Messages