Analog Bus Correction Constant (Test 46)

Analyzer warm-up time:
This procedure calibrates the analog bus by using three reference voltages: ground, +0.37 V
and +2.5 V. The calibration data is stored as correction constants in EE ROMs.
1. ress
PRESET 5 SYSTEM 5
2. Observe the analyzer for the results of the adjustment routine:
If the analyzer displays
If the analyzer displays
chapter.
3-8 Adjustments and Correction Constants
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
0 minutes.
ERVICE MENU TE T
, you have completed this procedure.
ABU Cor DONE
, refer to the \Digital Control Troubleshooting"
ABU Cor FAIL
EXECUTE TE T YE
46 5 x 5
.