HP 8753D Option 011 Service Manual page 137

Network analyzer
Hide thumbs Also See for 8753D Option 011:
Table of Contents

Advertisement

Step 4. Faulty Group Isolation
Use the following procedures only if you have read the previous sections in this chapter and
you think the problem is in the analyzer. These are simple procedures to verify the four
functional groups in sequence, and determine which group is faulty.
The four functional groups are:
power supplies
digital control
source
receiver
Descriptions of these groups are provided in Chapter 12, \Theory of Operation."
The checks in the following pages must be performed in the order presented. If one of the
procedures fails, it is an indication that the problem is in the functional group checked. Go
to the troubleshooting information for the indicated group, to isolate the problem to the
defective assembly.
Figure 4-3 illustrates the troubleshooting organization.
Figure TO4 here.
4-8 Start Troubleshooting Here
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Figure 4-3. Troubleshooting Organization