HP 8753D Option 011 Service Manual page 86

Network analyzer
Hide thumbs Also See for 8753D Option 011:
Table of Contents

Advertisement

ADC Offset Correction Constants (Test 52)
Analyzer warm-up time:
These correction constants improve the dynamic accuracy by shifting small signals to the most
linear part of the ADC quantizing curve.
1. ress
PRESET 5 SYSTEM 5
ote
This routine takes about three minutes.
2. Observe the analyzer for the results of the adjustment routine:
If the analyzer displays
If the analyzer displays
RAFT
3/21/106 15:11
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
0 minutes.
ERVICE MENU TE T
, you have completed this procedure.
ADC Ofs Cor DONE
, refer to the \Digital Control" chapter.
ADC Ofs Cor FAIL
EXECUTE TE T YE
52 5 x 5
Adjustments and Correction Constants 3-21
.