Node 24
2nd LO
Perform step A12, above, and then press MEAS
MENU
.
CW FREQ
5
This node counts the 2nd LO used by the sampler/mixer assemblies to produce the 2nd IF of
4 kHz. As you vary the frequency, the counter reading should change to values very close to
those indicated below:
Node 25
PL Ref (phase lock reference)
Perform step A12, above, and then press MEAS
.
MENU
CW FREQ
5
This node counts the reference signal used by the phase comparator circuit on the A11 phase
lock assembly. As you vary the frequency, the counter reading should change as indicated
below:
10-34 Service Key Menus and Error Messages
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
ANALOG IN 24
5
Frequency Entered
Counter Reading
0.3 to 1 MHz
frequency entered +4 kHz
1 to 16 MHz
not accurate
16 to 3,000 MHz
996 kHz
ANALOG IN 25
5
Frequency Entered Counter Reading
0.3 to 1 MHz
frequency entered
1 to 16 MHz
not accurate
16 to 3,000 MHz
1 MHz
x1
COUN ER: ANALOG BUS
5
5
x1
COUN ER: ANALOG BUS
5
5