CORREC ION CONS AN S NO S ORED
Error Number
store operation to the EEPROM was not successful. You must
3
change the position of the jumper on the 9 CPU assembly. Refer
to the \ 9 CC Jumper Position Procedure" in the \ djustments
and Correction Constants" chapter.
CORREC ION URNED OFF
Error Number
Critical parameters in your current instrument state do not match
66
the parameters for the calibration set, therefore correction has been
turned o . The critical instrument state parameters are sweep type,
start frequency, frequency span, and number of points.
CURREN PARAME ER NO IN CAL SE
Error Number
Correction is not valid for your selected measurement parameter.
64
Either change the measurement parameters or perform a new
calibration.
DEADLOCK
Error Number
fatal rmware error occurred before instrument preset completed.
111
Call your local Hewlett-Packard sales and service oce.
DEVICE: not on, not connect, wrong addrs
Error Number
The device at the selected address cannot be accessed by the
119
analyzer. Verify that the device is switched on, and check the
HP-IB connection between the analyzer and the device. Ensure
that the device address recognized by the analyzer matches the
HP-IB address set on the device itself.
DISK HARDWARE PROBLEM
Error Number
The disk drive is not responding correctly. Refer to the disk drive
39
operating manual.
RAFT
3/21/106 15:14
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Service Key Menus and Error Messages 10-43