rror Term Inspection
ote
If the correction is not active, press
1. Press
SERVI E MENU TESTS
SYSTEM
5
The analyzer copies the rst calibration measurement trace for the selected error term into
memory and then displays it. Table 11-2 lists the test numbers.
2. Press
and adjust the scale and reference to study the error term trace.
S ALE REF
5
3. Press
MARKER F TN
5
4. Compare the displayed measurement trace to the trace shown in the following \ rror
Term descriptions" section, and to previously measured data. If data is not available from
previous measurements, refer to the typical uncorrected performance specications listed in
Table 11-2.
5. Make a hardcopy of the measurement results:
a. Connect a printing or plotting peripheral to the analyzer.
b. Press
SYSTEM ONTROLLER SET ADDRESSES
LO AL
5
peripheral to verify that the HP-IB address is set correctly on the analyzer.
c. Press
SAVE/RE ALL
d. Press
MORE TITLE
DISPLAY
5
For detailed information on creating hardcopies, refer to \Printing, Plotting, and Saving
Measurement Results" in the
If Error Terms Seem Worse than Typical Values
1. Perform a system verication to verify that the system still conforms to specications.
2. If system verication fails, refer to \Start Troubleshooting Here."
Uncorrected Performance
The following table shows typical performance without error-correction. RF cables are not
used except as noted. Related error terms should be within these values.
11-6 Error Terms
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
32
x1
5
and use the marker functions to determine the error term magnitude.
and then choose either
5
and title each data trace so that you can identify it later.
P 8753D Option 011 Network Analyzer User's Guide
ORRE TION ON
AL
5
.
EXE UTE TEST
5
and select the appropriate
or
PRINT MONO HROME
.
.
PLOT
.
RAFT
3/21/106 15:14