HP 8753D Option 011 Service Manual page 111

Network analyzer
Hide thumbs Also See for 8753D Option 011:
Table of Contents

Advertisement

5. On the H 8753D Option 011, press
Option 006).
6. No adjustment is required when the spectrum analyzer measures 2.9 GHz ;5 kHz (or 5.9
GHz
5 kHz for Option 006). Otherwise, locate the A12 assembly (red extractors) and
adjust the VCXO ADJ for a spectrum analyzer center frequency measurement of
See Figure 3-19.
7. Replace the A12 assembly if you are unable to adjust the frequency as speci ed. Repeat
this adjustment test.
Figure VCXO here.
Figure 3-19. Location of the VCXO A J Adjustment
3-46 Adjustments and Correction Constants
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
CW FREQ
PRESET 5 MENU 5
(or
for
2.9 5 G/n 5
5.9 5 G/n 5
5 kHz.