Node 7
Log (log ampli er output detector)
Perform step A3, above, to set up a power sweep on the analog bus. Then press MEAS
ANALOG IN 7
x1
SCALE REF
5
5
Node 7 displays the output of a logger circuit in the ALC loop. The trace should be a linear
ramp. Absolute voltage level variations are normal. Flat segments indicate ALC saturation
and should not occur between 15 dBm and +10 dBm.
The proper waveform at node 7 indicates that the circuits in the A3 source ALC loop are
normal and the source is leveled.
Figure NODE7 here.
Node 8