Figure 10-2 Sampler And Sampling Oscillator Test Setup - HP 8560E Service Manual

Spectrum analyzers
Table of Contents

Advertisement

Synthesizer Section
Synthesizer Troubleshooting Section
23.Set the HP 8560E/EC to the following settings:
24.Set the HP 8560E/EC and HP 8340A/B frequencies to the
combinations listed in
the spectrum analyzer.
Figure 10-2
Sampler and Sampling Oscillator Test Setup
25.At each combination, the frequency counter should measure a
sampler IF as shown in
oscillator of the offset PLL tunes to the frequencies listed in the
table.) If the frequency counter does not read the indicated sampler
IF ±10 kHz, suspect the A15 RF assembly.
26.Reconnect W34 to A15U100J1 and W32 to A15J101.
27.The 1ST LO OUTPUT, located on the front panel, must be
terminated in 50 Ω. If the YTO unlocks only with certain center
frequency and span combinations, check that the termination is in
place.
28.Set the HP 8560E/EC
unlock conditions.
512
Frequency standard
Power level
Span ................................................................................. 0 Hz
Trigger .......................................................................... Single
Table 10-2 on page 513
CENTER FREQ
Table 10-2 on page
513. (The sampling
and
SPAN
EXT
−5 dBm
and press
on
SGL SWP
to generate the
Chapter 10