Measuring LCR Using Impedance Probe (HP E4916A with Options
001 and 010 Only)
Caution
^
Make sure that the test signal level is 0 dBm or lower when using the
probe. For more information, refer "EMC" in chapter 9 of Operation
Manual.
Required Equipment
H HP E4916A
Option 001 Impedance Probe Kit
Option 010 LCR Measurement Function
Connection
See
Connecting a Probe (HP E4916A Option 001 Only)" in
Chapter 3 for how to connect the impedance probe.
Selecting LCR Mode
■ Sequentially press
((blue)
) and
(Meas Time (lcr)I
At
this time,
LCE
appears at the lower left corner of the LCD to indicate that the LCR
mode has been selected.
Selecting Measurement Circuit
To use the impedance probe, you need to specify the probe as the
measurement circuit.
1- Use (i)=>) or (cr'll) to display CKT:nnrm . (nnnn shows the current
setting.)
Press CKT:nnnn to display the following:
CKT: PI PROBE BRIDGE NONE
3. Select PROBE.
Selecting Measurement Parameters
The HP E4916A can display up to 2 pairs of measurement parameters
or 4 parameters at the same time. The following lists parameters that
can be specified:
Parameter Specified
with Pril: and Pri2:
Parameter Specified
with Seel: and Sec2:
Z, Y, R, G, Gs, Cp, Ls, Lp
Oz, $y,
X, B, D, Q, Rs, Rp, G
1. Press
(Meas Printr)
to display the following:
Z Y R G Gs Cp Ls Lp
2. Select the desired parameter.
3. Next, press Sect:# to display the following:
#Z 0Y X B D Q Rs Rp G
5-22
Applications