System Trace Noise (Only For Analyzers With Option 006) - HP 8753E Service Manual

Network analyzer
Table of Contents

Advertisement

11. System Trace Noise (Only for Analyzers with
Option 006)
Specifications
Frequency Range Ratio System Trace Noise System Trace Noise
30
30
1 At + 5 dBm into test port, 3 lcHz IF bandwidth, and CW
Required Equipment
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 8120-4779
Perform this test to measure the system trace noise at designated CW
frequencies in both the A/R and B/R ratioed measurements.
1. Connect the equipment as shown in Figure 2-32.
System Verification and
Performance Tests
A/R
B/R
A/R
B/R
H P 8753E
N E T W O R K A N A L Y Z E R
Figure Z-32. System Trace Noise 'Ikst Setup
sweep.