Operation
End-expiratory volume measurement
End of operation
Only for stationary use!
End-expiratory volume measurement
(Special accessory)
To measure the end-expiratory tidal volume and end-
expiratory minute volume.
Cannot be combined with the PEEP valve.
Strictly follow the Instructions for Use of the
Volumeter 3000.
1
Clamp the Volumeter 3000 holder to the wall rail.
2
Screw the Volumeter 3000 to the holder.
3
crew the elbow connector to the Volumeter 3000.
4
Insert the hose adapter in the exhaust connector of
the ventilation valve.
5
Connect the ventilation valve and Volumeter 3000
with 1.5 m long ventilation hose.
End of operation
After disconnecting the patient:
6
Switch the main switch to 0.
If the Oxylog is supplied by 02 cylinder:
Fully close the cylinder valve.
If the Oxylog is supplied by the central mains supply:
Remove the gas supply connector.
10