Rs-232 Interface Connector - Thermo Scientific MaxQ SHKE416HP Operating And Maintenance Manual

High performance orbital shakers
Table of Contents

Advertisement

RS-232 Interface
Connector
Thermo Scientific
Table 1-4. Test Tube Racks and Adjustable-Angle Test Tube Rack Holders
Part Number
Description
950040
Test tube rack 10-13mm size
950060
Test tube rack 16-20mm size
600074
Test tube rack 21-25mm size
600075
Test tube rack 26-33mm size
600076
Adjustable-angle test tube rack holder w/rack 10-13mm
600077
Adjustable-angle test tube holder w/rack 16-20mm
600078
Adjustable-angle test tube holder w/rack 21-25mm
600079
Adjustable-angle test tube holder w/rack 26-30mm
600089
2 tier microplate rack
600090
3 tier microplate rack
#10-24 Pan head Phillips screws for mounting test tube holders to Thermo
194024
orbital shaker platforms
Pan head Phillips screws, washers and nuts for mounting test tube holders
185062
to Thermo 2568 and 2569 shaker baths
The MaxQ High Performance Orbital Shaker is equipped with an RS-232
Serial Communication Interface for the remote transmission of data. An
RJ-11 telephone style connector is located on the rear of the shaker. Refer
to Figure 1-14.
The data is "dumb terminal" formatted, which permits connecting to a
computer or a serial printer.
Three wires are used for the RS-232 interface:
1. Transmit data (/TXD) - pin 2
2. Receive data (/RXD) - pin 3
3. Signal ground (GND) - pin 7
The data format is:
Baud . . . . . . . 1200 (9600 baud w/ jumper at J2 on Main Control Board)
Data bits . . . . . . . .8 (7 bit ASCII with leading zero)
Start bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Stop bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Parity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .none